|
UV GUIDE
UK
Advances
in Reptile Lighting
A
resource for all reptile keepers
|
|
Using
the Solarmeter 6.2 UVB Meter
Instructions
for Use Solar
Recordings Recordings
in the vivarium
Setting
up a Test Bench Spread
Charts
Comparing
Meters Problems?
Feature:
Make Yourself a UVB Spread Chart
|
|

The
Solarmeter 6.2 UVB meter (fig.1) is easy to use and we have
found the readings to be consistent and reliable. We welcome
your recordings to add to our database.
Instructions
for Using your Solarmeter 6.2 with Reptile Lamps
1.
Open top flap of leather carry-case to uncover sensor.
2.
Aim the sensor at the light source. Do not touch
the sensor or hold it against the surface of the
lamp.
3.
Press the on/off button below the LCD display.
4.
Read LCD display, which gives the amount of UVB
reaching the sensor in microwatts per square centimetre.
5.
Using very slow, small movements, tip the meter
to scan for the best alignment - which will give
you the highest reading on the display. This is
the reading to record.
To
record the output of a lamp,
hold the meter with the sensor at a specific distance
from the lamp surface. We
suggest that if only one measurement is to be taken,
this is at the standard distance of 12
inches, or 30 centimetres.
Lamps of the same brand anywhere in the world may
only be compared easily if all recordings are made
at the same distance from the lamp. We welcome your
recordings to add to our database.
12" is a useful "standard" distance
since it is often stated to be the maximum
distance at which a standard reptile UVB fluorescent
tube should be used, and several mercury vapour
lamps are designed with this as the minimum
recommended
distance from the reptile.
Some more powerful lamps have minimum recommended
basking distances greater than 12". For these
lamps we suggest that at the very least, a recording
be made at their "recommended" minimum
distance, as well.
To
find out the UVB your reptiles are receiving,
hold the meter in the basking spot, so that the
sensor is at the level of the reptile's back, aimed
directly at the UV lamp.
|
A
little more detail...
The
Solarmeter Model 6.2 UVB
The meter we used in all our tests, the Solarmeter 6.2 from
Solartech, Inc. (www.solarmeter.com)
is claimed to be the most accurate hand held UVB meter on
the market. This very instrument has been used in current
research into vitamin D3 synthesis at the Vitamin D, Skin
and Bone Laboratory at Boston University Medical Center.19
The device is small, fitting easily in one hand, and is
powered by an ordinary 9v battery. The sensor is at the
top of the meter, and to take a reading, the operator simply
aims the meter directly at the UVB source and presses a
button on the front of the unit. The amount of UVB reaching
the sensor at that moment is displayed on the LCD panel
in microwatts per square centimetre (uW/cm²). If, for example,
the reading was 20uW/cm², this would mean that a square
centimetre of reptile skin would be getting 20 microwatts
of UVB at that point.
The sensor is extremely sensitive, with minute changes in
the angle between sensor and source causing variations in
the readings. Logically, the highest readings will be seen
when the sensor is most accurately aligned in the beam of
light; when taking hand-held recordings, it is adviseable
to "scan" for the best alignment with very slow,
very small movements. With practice, lining up the
sensor can be done fairly quickly and very accurately. When
taking recordings, we always use the highest readings obtained
consistently (i.e. on at least 2 to 3 "scans")
on each occasion.
Although it measures the complete UVB range (280 to 320nm)
the Solarmeter's peak sensitivity is between 290-300nm,
right in the middle of the range of wavelengths responsible
for D3 synthesis, which makes it ideal for checking UVB
reptile lamps.
Our Solarmeters were imported from the USA. However, there
are now several Solarmeter distributors in the UK, and Zoo
Med have just brought out a very similar meter, also made
by Solartech, Inc., in the UK, which can be obtained from
British Zoo Med stockists. (see our links
page for details of all distributors and suppliers)
Measuring UVB with
the new meters
What we have been able to do, with our hand-held radiometers,
is to look at the entire UVB output of the sun, and of any
light source, between 280 and 320nm of the electromagnetic
spectrum and investigate in detail the initial output of
a new lamp; the decay in the output of a lamp with time;
and the characteristics of the UVB "beam" for each type
of lamp - how far the UVB light extends from the lamp and
in what directions.
Between
us we have taken many thousands of readings both outdoors
in natural sunlight and shade, and indoors from a wide range
of UV products.
There's
more detail about our project, the accuracy of our readings
and comparisons with other work in this field in the Introduction
to the 2005 Lighting Survey
The
sections below describe the methods of recording we have
found the most useful when using the meter outdoors (for
solar recording), indoors in the vivarium, and on a simple
"test bench".
The
page concludes with a short section on comparing different
UVB meters, and a discussion of one or two problems which
might be encountered when using the Solarmeter 6.2, along
with their solutions.
|
|
Solar
Recordings
The
readings collected for our studies on UV
light in nature are mainly direct
readings, for which the meter is aimed directly at
the sun. These give a picture of the maximum UVB which a
reptile could obtain if it was fully exposed to sunshine
at that location.
When
taking solar recordings, never
look directly at the sun. It is easiest to align
the meter whilst facing the sun. Hold the meter out at a
comfortable distance at just below shoulder level and "scan"
carefully by tipping the meter towards the sun, searching
for the highest reading.
Readings
need to be taken out in the open, where trees or buildings
do not block a clear view of the sky. The sensor's field
of view is a full 180° with a cosine response curve - this
means that although it is most sensitive to whatever is
directly in line with it (in this case, the sun), it will
include diffused UVB from the sky around the sun in the
reading. If this "sky" includes the dark silhouette
of a building or tree, a lower reading will be obtained.
Conversely, if there are large reflective surfaces such
as water or snow beneath the sky, reflected and scattered
UV from these may result in a higher meter reading.
Close
to sunrise and sunset, readings from the sky above the sun
may be slightly higher than those from the sun itself. This
is because light from closer to the horizon must pass through
a thicker layer of atmosphere. Water vapour in the air scatters
light of shorter wavelengths more strongly than longer wavelengths;
close to the horizon this thicker layer scatters enough
short-wavelength light to remove most of the ultraviolet
and blue. (This is why, of course, the rising or setting
sun appears only red or orange.)
If taking
a series of readings throughout a single day, ideally all
the readings should be taken from the same spot, so that
the readings are comparable one with another. It is also
important to record the exact time of each recording. As
the sun rises in the sky, the reduction in the thickness
of atmosphere through which the light is passing causes
an extremely rapid increase in UVB. We have recorded changes
of a microwatt a minute. For example, between 7.30am and
8am on June 10th 2005, in Wales, UK, under a clear sky,
the reading climbed from 40 to 70uW/cm². By 8.30am it was
105uW/cm² and by 9am it was 135uW/cm².
When
taking solar recordings, it's useful to note the weather
conditions; in particular, if the sun is shining, is it
in a clear, deep blue sky or is there high cloud or haze?
The increase in water vapour in the sky due to these will
reduce the UVB levels reaching the ground.
If
there is total cloud cover, direct solar recordings may
not be possible. So-called "global" recordings
usually give the highest readings with 100% overcast skies.
To obtain these, the meter is held vertically, to record
from the zenith (directly above the observer).
When
recording for a database, you may also wish to record the
location (latitude and altitude) at which the recording
was made. If your country is using Daylight Saving Time
this too needs adding to your report; and remember, if so,
the sun will reach its highest point in the sky, and UVB
its maximum level, around 1.00pm Local Time rather than
at noon.
Fig.4
shows recordings made during the course of one clear day
in August in Wales, UK. Faint high cloud and haze in front
of the sun has caused slight "dips" in the graph
at various times but the pattern is distinctive. More details
on solar recordings, and the ongoing Solar Recording Project,
are in our feature UV
light in nature and in the files section of the UVB_Meter_Owners
internet mailing group. We welcome your solar recordings
from anywhere on the planet, to add to our database.
|
|
Recordings
in the Vivarium
Safety
Precaution: Always wear eye protection when studying UV
lamps.
However
careful you are, it is very difficult to take recordings
without looking directly at the lamp, especially when measuring
the distance from the lamp to the meter. Look for glasses
which state clearly that they offer 100% protection from
UV light.
Whereas
some fluorescent tubes, for example, emit (at 12" distance)
less UV light than is reflected by sunlit trees and
grass, and are thus unlikely to cause harm, some mercury
vapour lamps, at 12" distance, emit as much UVB as
the sun at the centre of their beam.
Intense
visible light may also damage your eyes. Even when wearing
glasses, avoid staring directly into any very bright lamp,
whether or not it emits UV, at close range.
Taking
readings
The
simplest recordings are taken using just a meter and a steel
tape measure, ruler or measuring stick. The end of the tape
measure is positioned at the surface of the tube or face
of the lamp, and the meter is positioned so that the sensor
is at the required distance from the light.
Because
the tape measure or ruler will inevitably block some light
from the sensor, it must be moved away before the reading
is taken. This manoeuvre takes a little practice, as the
meter must remain at the correct distance while the tape
measure is removed.
With
fluorescent tubes, the highest
readings will be found when the meter is positioned perpendicular
to the axis of the tube, half-way along its length.
With
lamps and bulbs, the beam is
usually focused by internal reflectors and although, in
general, the highest readings are found directly in front
of the lamp face, this cannot be relied upon. Often, the
beam is at a slight angle from the perpendicular owing to
minute discrepancies in the alignment of the arc tube, glass
envelope, etc. The path of the UVB beam usually closely
follows that of the visible light and this can be used as
a first step in aligning the meter within the beam; thereafter,
careful scanning (in the same way as previously described
for a solar reading) will locate the "hot-spot"
at the chosen distance.
Fig.5
shows a steel tape measure being used to take a reading
from a fluorescent tube fitted with a reflector. First the
meter is positioned at the required distance (in this example,
6 inches). Next, the meter is switched on and moved around
a little, at that distance, whilst scanning for the maximum
reading (i.e., aligning the meter). Then, the tape measure
is moved out of the way, a final alignment made and the
reading taken. In this example, the reading increased from
102uW/cm² to 105uW/cm² with the removal of the tape measure.
Sets
of readings may be taken from each lamp tested, at increasing
distances from the lamp, to give an indication of how far
the UVB penetrates into the vivarium. If readings are repeated
at intervals (for example, once a month), the decay of the
lamp throughout its lifespan may also be recorded.
If there
is more than one UV source in a vivarium, don't forget to
switch off the one you are not recording from, or it will
affect your results.
Watch
out for metal reflections in the vivarium. UVB is hardly
reflected at all from glass or perspex, which absorbs the
rays. Glass mirrors likewise reflect almost no UVB. However,
bare metal surfaces such as polished or brushed aluminium
reflect UVB strongly. You may obtain false high readings
from a lamp if the sensor picks up UV from the lamp's reflection
in a metal surface.
|
|
Setting
Up a Test Bench
If
you want to take your study further, you may wish to set
up a simple test bench for the rapid, accurate measurement
of the UVB output of your lamps. New lamps can be burned-in
whilst on the bench, making it easy to repeat daily measurements
precisely. Lamps on long-term testing can be temporarily
removed from the vivarium and placed on the bench for individual
tests, as well.
A simple
test bench for fluorescent tubes and compact lamps is easily
constructed on a table top or work surface. Fig. 6 shows
recordings being made from a fluorescent tube. Dark coloured
card has been placed on, and behind, the work surface to
reduce reflections.
A measuring line has been drawn on the card and the tube
mounted vertically, so that the zero for the measuring line
is at the surface of the lamp. Compact lamps may be place
in simple lamp-holders positioned at the end of a measuring
line, in the same way.
To
make life easy and eliminate the need for a tape measure,
the UVB meter has been temporarily attached to a set square
such that a pointer fixed to the base of the square, resting
on the measuring line, indicates the position of the sensor
directly above it. The distance from the tube to the meter
sensor can thus be measured accurately without a tape measure.
The meter is held steadily against the set-square while
it is moved along the measuring line, enabling rapid recording
of a set of measurements with no errors due to hand-shake.
Mercury
vapour lamps, because of their directional beam, are more
conveniently tested whilst hanging pendant-fashion. If the
lamp-holder is secured to a ceiling or a horizontal beam
close to a wall, the measuring line may be drawn on the
wall, or on card affixed to the wall (Fig. 7.)
|
|
Spread
Charts
A
further development to the recording of the output of a
lamp in this simple linear style, is the production of a
"spread chart". This is a two-dimensional map
of the shape of the beam, constructed by plotting the UVB
gradient under the lamp. Spread charts for each type of
lamp are featured in the 2005 Lighting Survey reports.
To construct
a spread chart is not difficult, but it is time-consuming.
If you would like to make one yourself, full details may
be found in our article: Make
Yourself a UVB Spread Chart
|
|
Are
all meters equal?
All
our UVB data has been compiled from readings taken with
the Solarmeter
6.2 UVB meter. These
Solarmeters are individually calibrated by the manufacturer,
by reading transfer from a standard meter calibrated to
NIST standards. They are claimed to be accurate to at least
±10%. We have tested a number of meters together,
and have found that individual meters are consistent with
each other in the readings they give from the same light
source (as demonstrated in fig. 9 with models of the Solarmeter
6.2 and the ZooMed UVB meter,
made with the same components by the same company). We therefore
believe that it is valid to collate, combine and compare
data from all Solarmeter 6.2 UVB meters and we are interested
in collecting data from owners of these meters worldwide.
For
technical reasons, however, it is
not possible
to combine or compare raw data from two
different brands of radiometer, or even models from
the same manufacturer with different specifications. Experiments
conducted by Gehrmann et al (see references 18
and 19
on our reference page) have shown that the most accurate
UV radiometers from different manufacturers may give very
different readings from the same light source. Reasons for
this may include differences in the response to different
wavelengths by different components in the sensors, or in
their calibration. Gehrmann and his team used measurements
of the in-vitro production of vitamin D3 under UV light
to devise conversion factors for the readings from four
different meters. (Such work is outside the scope of this
project.)
|
|
Problems?
Low
Battery
The
Solarmeter runs off a standard 9-volt DC battery which may,
according to the manufacturer, last for about two years
of "normal" use. Battery operation voltage is
9V down to 6.5V. Below 6.5V the LCD numbers will begin to
dim, indicating the need for battery replacement. The readings
remain accurate until the numbers begin to dim.
High Humidity and Heat
The
meters are sensitive to excessive humidity and heat. If
the meter is stored or used in a very hot or humid situation,
there can be a temporary false elevation of the reading
by several microwatts. This is easily detected, as the meter
will not read "zero" with the cover over the sensor,
but give a reading of a few microwatts instead. Allowing
the meter to cool or dry out will resolve the problem. In
very humid environments, the manufacturer suggests storing
the meter in a plastic bag with a sachet of silica gel.
When
testing a high-wattage lamp, at close range the meter may
be affected by the heat. Our tests using two 100 watt mercury
vapour lamps showed that even as close as 4" from the
lamp surface, provided the readings are taken within 1-2
minutes there is little or no effect upon the reading. Longer
exposure to heat does cause a slow artificial rise in the
readings as the front of the meter gets hot. A change in
temperature from around 21-22°C, to around 44-50°C
produced a change in readings from 1 - 6uW/cm2 in a set
of 4 experiments- a change of between 2-6%.
It is
worth bearing in mind that readings taken close to lamps
should therefore be taken fairly rapidly, within maybe a
half a minute or so, as holding a meter close to a hot bulb
for any length of time will cause a slow false rise in the
readings as the sensor is heated.
The
meter sensor should not be touched or held against the
surface of a lamp. Contact with any solid object can
produce minute electrical fields which create brief false
readings. More seriously, contact with the extremely hot
glass of some incandescent or mercury vapour lamps could
easily damage the sensor.
|
|
© 2006 UVGuide.co.uk
|